Purpose of Equipment
For long term or short term storage of various vaccine requiring low temperature (below ambient) to ensure appropriate and quality cold-chain especially in areas susceptible to power failures.
System Features
• Microprocessor temperature controller
• During power failure, battery back-up system supplies power to the controller for at least 36 hours
• Non-volatile memory ensures all set values remain intact in case of power failure
• Lock-temperature function protects against tampering of set temperature/values
• Off-set function for calibration use. Important for ISO/Calibration requirements.
Optional Items
• Data Loggers
• Temperature Monitoring System
| MODEL | VR-70 |
|---|---|
| Capacity | 74 litres |
| Cabinet construction | Powder-coated steel plate exterior, stainless steel interior |
| Cabinet insulation | CFC-free rigid polyurethane, foam-in place insulation |
| Working temperature range | 2°C to 8°C. Preset at 4°C |
| Hold over time during power cut | Better than 12 hours |
| Alarms | Audio and visual alarms with battery back-up system. Supplied with remote alarm contact as standard |
| Alarm conditions | High-Alarm, Low-Alarm, A.C. Failure, Low Battery, Door Open and Clean Filter |
| Alarm set points | Preset at 0°C for under temperature and 10°C for over temperature |
| Safety features | Supplied with RCCB and high-low pressure switches as standard |
| Cooling capacity at 5°C | 370W |
| Door type | Dual-pane heated single glass door. Supplied with door lock as standard. |
| Shelf type | Adjustable stainless steel shelves |
| No. of shelves supplied | 2 units |
| Shelf dimensions (mm) | 480W x 450D x 10H |
| Interior dimensions (mm) | 365W x 430D x 473H |
| Exterior dimensions (mm) | 600W x 615D x 985H |
| Electrical | 220-240 VAC, 50 Hz, 1.6A |
| Weight | 63kg |