Suitable for testing of Small Parts (in Laboratories)
These bench top chambers are equipped with a high precision PID controller and well-prepared safety devices such as no fuse breaker, overheat protector, and refrigeration overload relay.
Model with Programmable Control Function
The model MHGB 120CR(L) chambers offer programmable control function. The capacity of memory is 10 programs, 100 steps in total. The setting time per step is up to 99 hours 59 minutes in the resolution of one minute. Each program can be repeated for 999 times.
| Model | HRMB-120CR(Q) |
|---|---|
| Controller | Fix point PID control LCD display 0.1℃ 1% resolution |
| Temperature range | -20℃~100℃ (150℃) |
| Humidity range | 30 ~ 95% RH |
| Temp. & Humid. Constancy | ±0.5℃ ±2.5% RH |
| Temp. & Humid Uniformity | ±7℃ (Below 80℃) ; ±1.0℃ (80℃ ~ 100℃) ; ±1.5℃ (101℃ ~ 150℃) ; ±3% RH |
| Heating up time | -20℃ to 100℃ within 40 min |
| Pull down time | 20 ℃ ~ -20 ℃ within 60 min |
| Interior dimensions cm | W50/H60/D40 |
| Exterior dimensions cm | W120/H97.5/D98.2 |
| Interior & Exterior material | Stainless steel plate (SUS 304) |
| Insulation | Rigid Polyurethane Foam and Glass Wool |
| Refrigeration system | Single Stage Refrigeration System; Hermetic Compressor |
| Safety devices | Refrigerator overload relay; Refrigerant high pressure switch; Porcelain fuse; Lack of water protector; Magnetic switch; Boll-dry protector; Alarm |
| Accessories | Cable port (50 mm); Shelves (freely adjustable)x2 pcs; Wet-bulb wick x 5 pcs; Power supply cord 2.5 cm |
| Weight (kg) | 236 |
| Max power consumption (KVA) | 3.2 |
| Power | 12W+G AC220V ±10% 50/60 Hz |
| Notes : According to the standard of JTM K03-1998, the SENSOR should be located at a distance of 1/6 of the interior dimension from the interior chamber wall. | |
| The above specifications are for reference only; No further notice will be issued and take the specification in the quotation as standard if any change occurs. | |